Thermometer
Model:ET-6803A
1. Specifications:
1). With the function of temperature measurement and output alarm.
2). Supporting 8 kinds of thermocouple: K,E,J,T,N,B,R,S.
3). C and F display.
4). Dual Display: It can display the input temperature Max./ Min/mV value.
5). Seting alarm when it is higher or lower than alarm setting point; and set the alarm cadent value if phonate or not.
6). With the Memory Function: Support manual memory about 2500pcs temperature points; support auto memory about 6000pcs temperature points, memory space can be adjusted, the longest memory time is about 99 hours.
Thermocouple Type |
Input Range |
Resolution |
Accuracy |
R |
-40~1760C |
1C |
-40~500C: ¡À0.5%+2C)
500C~1760C: ¡À(0.4%+2C) |
S |
-20~1760C |
1C |
B |
400~1800C |
1C |
400C~800C: ¡À(0.5%+2C)
800C~1000C: ¡À(0.4%+2C)
1000C~1800C:¡À(0.3%+2C) |
E |
-200~1000C |
0.1C |
-200C~0C: ¡À(0.3%+1C)
0C~500C: ¡À(0.3%+2C) |
K |
-200~1370C |
0.1C |
-200~0C: ¡À(0.3%+2C)
0C~800C: ¡À(0.3%+1C)
800C~1370C: ¡À(0.5%+2C) |
J |
-200~1200C |
0.1C |
-200~0C: ¡À(0.3%+1C)
0C~1200C: ¡À(0.3%+2C) |
T |
-200~400C |
0.1C |
-200~0C: ¡À(0.3%+1C)
0C~400C: ¡À(0.3%+2C) |
N |
-200~1300C |
0.1C |
-200~0C: &n, bsp; ¡À(0.5%+2C)
0C~1300C: ¡À(0.3%+2C) | |